1) Select PALETTE.
.
Indicator. This table will show below-range pixels:
in blue and saturation in red.
3
Select CONT. and ad
ust Detector Gain until few
or no red pixels remain. Adjust Amplifier Offset until
most blue pixels are just gone. Each channel must
be adjusted independently (choose SPLIT XY to
see each channel separately).
4) Select No Palette to return to normal channel
display.
NOTE: If needed, you can use a higher
NOTE: In multi track, it’s best done one channel
at a time (only one laser line/one detector
.
,
but also significantly increase noise. You may
also need to increase/decrease laser power.
2/27/2008
15
ac
ve
.
sw
ncrease spee
an
re
uce
bleaching. Select/Deselect individual tracks in
multi track mode on the CONFIG panel.