START-UP
Carl Zeiss Installing standard components Axio Lab.A1
54 430037-7144-001 04/2013
3.1.8 Mounting rotary stage Pol
3.1.8.1 Removing rotary stage pole
x Loosen screw cap (Fig. 3-12/6) from the spring
housing (about three rotations).
x Press rotary stage Pol (Fig. 3-12/4) to the front
against spring-loaded pin (Fig. 3-12/7), lift it off
the stage carrier (Fig. 3-12/5) from the back and
remove it upwards.
x Re-tighten screw cap (Fig. 3-12/6).
3.1.8.2 Attaching rotary stage Pol
x Where necessary, loosen screw cap (Fig. 3-12/6)
of spring housing with approx. three rotations.
x Place rotary stage Pol with the groove of the
dovetail (stage bottom) on spring-loaded pin
(Fig. 3-12/7).
x Attach the rotary stage with the clamp screw
(Fig. 3-12/8) pointing to the front right.
x Press the rotary stage Pol to the front against
the spring-loaded pin and lower it towards the
back into the stage carrier (Fig. 3-12/5), then
release it.
x Re-tighten screw cap (Fig. 3-12/6).
The rotary stage must be mounted so that the vernier scale is on the left side and clamp is on
the right side.
3.1.8.3 Dismantling the detachable specimen guide and mounting stage clips
x Loosen the knurled screw (Fig. 3-12/1) on the specimen guide Pol. Withdraw the specimen guide Pol
(Fig. 3-12/2) upwards.
x Insert the stage clips (Fig. 3-12/9) into the holes provided.
Fig. 3-12 Replacing snap-in rotary stage Pol,
detachable specimen guide Pol and
stage clips