START-UP
Carl Zeiss Installing standard components Axio Lab.A1
58 430037-7144-001 04/2013
3.1.10 Installing or replacing 35 W
halogen bulb or 3 W white
light LED lamp
If desired, Axio Lab.A1 transmitted light stands can
be equipped with a 3 W white light LED lamp with
spectrum daylight or warm light.
To insert or replace the halogen/LED lamp, proceed
as follows:
x Switch off the microscope, remove the power
cord on the microscope and allow it to cool
down at least 15 min.
x Press the clamping jaw (Fig. 3-16/1) on the
cover (Fig. 3-16/2) downwards. Swing the cover
down, remove from the retaining channel
(Fig. 3-17/4) on the stand and set aside.
x Remove the lamp plug (Fig. 3-17/2) from the
lighting unit (Fig. 3-17/3).
x Press the loops (Fig. 3-18/1) on the securing clip
(Fig. 3-17/1) of the lamp holder together and
swing it out to the front.
x When changing the lamp remove the lighting unit
(Fig. 3-17/3).
x Position the new lighting unit/LED lamp
(Fig. 3-18/1) with the lower front edge between
the contact surface and securing clip. Set the
light fitting on the securing clip.
x Lift the securing clips (Fig. 3-18/2) of the lamp
holder with the lighting unit until the latter is
completely enclosed in the lamp holder. In
doing so, slightly press the ends of the securing
clips together and guide past the two upper
retaining elements. Release the pressure until
the clamping bracket opens and engages on
both sides in the retaining elements.
Fig. 3-16 Removing the cover
Fig. 3-17 Removing the LED lamp
Fig. 3-18 Inserting the LED lamp