Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
2Ć10
2.4 Illumination and contrasting techniques
The description/application of illumination/contrasting techniques is based on the following
microscope settings:
• Axiovert 25 microscope is switched on.
• The "Illuminance" control (1Ć6ā/ā25) is in center position.
• The "beam path" switch (1Ć6ā/ā30) is in the position for observation.
2.4.1 Use of condensers
Condenser 0.2
0.4 0.55
Free distance between
condenser and stage
surface
90 mm 53 mm 31 mm
plus 6 mm height
travel
Objective magnification 5x, 10x, 20x 5x, 10x, 20x, 32x, 40x 5x, 10x, 20x, 32x, 40x
Technique brightfield
phase contrast
inclined illumination
brightfield
phase contrast
VAREL contrast
(up to 32x)
inclined illumination
brightfield
phase contrast
VAREL contrast
inclined illumination
• To fully illuminate object fields even when lowĆpower objectives (< 5x) are used, push condenser
(2Ć14/3) out of its work position (dashed line in Fig. 2Ć15).
• When particularly high culture vessels are used, this position of the condenser is also required
in order to increase the free working distance to approx. 190 mm.
NOTE Only simple general illumination is obtained if no condenser is used!
• When using the condenser again for illumination, slide it exactly up to the stop position.
NOTE Loosening the three screws (2Ć15/7) of the condenser changer permits the condenser
0.4 to be replaced with the condenser 0.2 or 0.55.