Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
3Ć6
• Use the SW 3 screwdriver to loosen the clamping screw on the heatsink and remove the used
lamp.
 The cable on the heatsink must not be removed.
• Hold the new lamp on the marked metal base and insert it in the heatsink in such a way that its
reflecting part is at the bottom when inserted in the mount. The reflecting meltĆoff part of the
lamp must point to the side (to avoid interference with the image).
• Carefully tighten the clamping screw on the heatsink.
• Avoid fingerprints on the glass parts of the lamp; if required, remove fingerprints immediately.
• Press down spring lever and insert the lamp in the mount while holding it on the heatsink.
• Slowly release pressure from spring lever and let go of the heatsink.
 The heatsink must be aligned parallel to the lamp housing. The position of the
heatsink can be corrected by pressing down the spring lever and turning the heatsink.
• Attach the lamp housing to the lamp base and tighten screw (3Ć4ā/ā1).
• Attach the microscope illuminator to the microscope stand.
• Note down the number indicated on the operating hours counter of the power supply unit. The
HBO lamp must be replaced when the nominal life of 100 hours has been reached.