B 40Ć012 e 08/99 2Ć15
NOTE • Shifting the VAREL illumination to outside the pupil corresponds to oneĆsided
darkfield illumination.
• Shifting the VAREL illumination between the Ph and VAREL rings of the objective
corresponds to inclined brightfield illumination.
• The middle engaging position of the Phā/āHā/āVar slider permits brightfield
illumination.
• The left engaging position of the Phā/āHā/āVar slider permits the fast change from
VAREL to Ph contrast.
Figure 2Ć20 Pupil images in VAREL contrast
Objective
VAREL contrast
Inclined brightfield
illumination
OneĆsided darkfield
illumination
Ph ring VAREL ring VAREL diaphragm
Preparing the Var/H/Var slider for phase contrast
• Remove the attenuation filter (2Ć21ā/ā3) from
the center position of the slider (remove
spring and filter).
• Insert the ring diaphragm mount (2Ć21ā/ā1)
(contained in the delivery package of the
slider).
If required, loosen centering screws (2Ć21ā/ā2)
using the key.
• Insert ring diaphragm in ring diaphragm
mount and center the ring diaphragm with
regard to the phase ring of the objective
using the centering screws (one centering
screw at each side of the slider).
Figure 2Ć21Varā/āHā/āVar slider