208 Operation
G-30-1458-en OPMI® Pentero® Software Release 2.20 / 2.21 Issue 11.1
Printed on 18. 02. 2009
What to do in an emergency
Illumination failure - changing the xenon lamp
• Open flap (3) as far as it will go.
Lever (1) for changing the lamp moves out.
• Pull out lever (1) as far as it will go, without using too much force.
If you do not properly pull the lever, the lamp is not changed. Pull the
lever again to make the lamp snap in.
• Close the flap again.
• Order a new xenon lamp. Cat. No. (2) is shown on the lamp module.
Note:
– Only change the used lamp if replacement is available on site (for the
lamp change, see page 222).
Always operate the illumination system with both xenon lamps. If one
of the lamp fails, switch to the second lamp. Leave the failed lamp in
the lamp housing until it is replaced by a new lamp.
– Every time you switch on the system, it indicates that you are using
the backup lamp, until the used lamp is replaced.
Kommentare zu diesen Handbüchern