354 DICOM (option)
G-30-1458-en OPMI® Pentero® Software Release 2.20 / 2.21 Issue 11.1
Printed on 18. 02. 2009
Caution:
It is the user's responsibility to ensure that the system environment is
free from viruses.
The network connector must be adequately contact-protected, e.g.
made of plastic material.
The cable and connector of the network connection must at least com-
ply with Cat-5e EIA/TIA-568A-5, i.e. the more recent Class D values
from ISO/IEC 11801:2002 or EN 50173-I:2002.
Do not use loaded images and videos for diagnostic purposes.
Always enter the patient ID to ensure that data to be stored in a DI-
COM archive are not assigned inadvertently to the wrong patient.
If you change a patient ID imported via DICOM, it cannot be guaran-
teed that the data is assigned to the correct patient when it is stored in
the archive.
Select the correct PACS.
Check the connection configuration with the PACS; otherwise there is
the risk of data being lost.
Correctly implement the TCP/IP configuration.
Perform a connection test.
Use the DICOM standard.
Export to CD/DVD or USB stick is only enabled after authentication by
the user and password entry.
In addition, there is the possibility of anonymizing DICOM data.
Warning!
Data transmitted by the system into the data network or data available in
the data network risk to be corrupted or transmitted incompletely. There-
fore, no liability can be accepted for the correctness of the data.
The operator of the data network is responsible for compliance with the
legal requirements regarding data security and for the protection of per-
sonal rights.