234 Technical data
G-30-1458-en OPMI® Pentero® Software Release 2.20 / 2.21 Issue 11.1
Printed on 18. 02. 2009
Line connection Only connect the suspension system to power
outlets which are provided with a properly con-
nected protective ground conductor.
Rated voltage 115 VAC (100...125 VAC ± 10%)
230 VAC (220...240 VAC ± 10%)
Current consump-
tion
115 VAC max. 12 A
230 VAC max. 6 A
Rated frequency 50...60 Hz
Fuses Automatic circuit breaker
Electrical
outlets
Power outlet 115/230 V, max. 2 A
AUX port for an external signal of a maximum
of 24 V / 0.5 A.
Electrical
standard
Complying with IEC 60601-1 / EN 60601-1; UL
60601-1;
CAN/CSA-C22.2 No. 601.1-M90
Protection class I, degree of protection IPXO
Type B equipment
Product classifica-
tion
As per Directive 93/42/EEC, Annex IX:
Class I
With intraoperative fluorescence: Class a
DICOM (option): IT network connector, class A,
as per EN 60601-1
Approval
EMC
requirements The system meets the EMC requirements of
IEC 60601-1-2.
The system meets the RFI requirements of Class
A (hospital grade).
CE labeling The system meets the essential requirements
stipulated in Annex I to the
93/42/EEC directive governing medical devices.
The system is labeled with:
AUX port
View of connector
side
1
2
3
1
2
3
176164
0297