306 FLOW 800 (option)
G-30-1458-en OPMI® Pentero® Software Release 2.20 / 2.21 Issue 11.1
Printed on 18. 02. 2009
Description of FLOW 800
Starting the FLOW 800 processing mode
Caution:
To obtain optimum FLOW 800 data, avoid changing the following param-
eters during data recording:
- Zoom factor
- Working distance/focus
- Light intensity
- Position of the OPMI stand, i.e. unlocking of brakes or joystick-
controlled movement
- Position of the light field aperture
The following influences are corrected automatically:
- Camera gain adjustment (via AGC or manually)
- Slight
wobbling of the microscope
(e.g. during operation of the handgrip buttons)
1 In the INFRARED 800 Setup menu:
The INFRARED 800 Setup menu includes the "FLOW 800"
processing mode which permits activation and deactivation of FLOW
800 processing. This requires a specific license. If this license is not
available, "FLOW 800" is disabled.
The FLOW 800 processing phase can be started at the end of the Re-
play phase by pressing the fluorescence button if the FLOW 800 pro-
cessing mode has been activated in the INFRARED 800 SETUP
menu.
2 In the Patient File menu:
You can also activate FLOW 800 processing from the Patient File
menu for previously recorded IR800 data, provided that this data is
processable, i.e. that it is marked with "A".
In the VIDEO CLIPS tab, select a video from the list of all videos of a
patient. The thumbnail of the video must be marked with "
A" (recorded
with analysis data) on the bottom right. The FLOW 800 button on this
page is enabled, and the processing mode can be started by pressing
this button.
The data is checked before the INFRARED 800 video is displayed in
the processing mode.
The data is not optimal if OPMI settings were changed (e.g.
change of focus) during recording of the INFRARED 800 data. In
this case, a message box with a note to this effect appears before
the data is displayed.
!
Recorded FLOW 800 processing data not optimal!
Changes in OPMI settings occurred during recording!
Please note instructions in user manual!
Note!
OK