Zeiss LSM 510 META Spezifikationen Seite 151

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 330
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 150
OPERATION
LSM 510 Image Optimization
B 40-051 e 07/98 4-93
4.7.3 Scan speed
The signal-to-noise ratio can be substantially
improved by reducing the scanning speed to an
acceptable level and averaging over several scans
(i.e. with an Average Number greater than 1 in
the "Scan Control" window).
Use the Scan Speed slider on the Speed panel
to set the slowest acceptable scanning speed.
The corresponding pixel scanning time (Pixel
Time) and the total scanning time (Scan
Time) are shown in the dialog box.
In the Number text box of the Pixel Depth,
Scan Direction & Scan Average panel enter
the number of measurements to be averaged.
Image optimization can be effected
much faster if you select a smaller
frame, since less data have to be
processed.
Fig. 4-102
Seitenansicht 150
1 2 ... 146 147 148 149 150 151 152 153 154 155 156 ... 329 330

Kommentare zu diesen Handbüchern

Keine Kommentare