INTRODUCTION TO LASER SCANNING MICROSCOPY
Functions LSM 510
5-58 B 40-051 e 07/98
Use the parameter Reflection to control the ratio of diffuse and reflective brightness components, i.e.,
the overall basic brightness compared with the highlights. When the value of Reflection is low, the
highlights predominate; when the values are high, the region appears to be uniformly illuminated and
the highlights are not so pronounced. When Auto Update is selected, the reconstruction is updated
automatically whenever a parameter is modified (except Input, Output, or Number of Views). Show
Cube defines whether a wire frame cube is shown in the Display window or not.
Application
This method can be applied, if the structures in the Input sequence can be segmented by grey value
thresholding. Because the gradient is calculated for every pixel, the Output appears in very fine detail.
Noisy Input sequences must be smoothed (function Smooth) before rendering, otherwise the surface
appears rough.
Parameters:
Input Input image sequence
Output Resulting image sequence
Number of Views Number of reconstructions to be calculated
Angle X Angle of rotation on the X-axis, start position
Angle Y Angle of rotation on the Y-axis, start position
Angle Z Angle of rotation on the Z-axis, start position
Channel All - The following parameters are valid for all channels
X - The following parameters are valid for the selected channel only
Grey Low Low grey value threshold of the region to be displayed
Grey High High grey value threshold of the region to be displayed
Aperture Measure of the extent of the highlights
Reflection Weight of the defuse brightness components in comparison to the
highlights
Auto Update 0 - Function execution is performed on OK or Apply
1 - Function execution for the current angle is performed on any parameter
change
Show Cube 0 - The wire frame cube is not shown
1 - The wire frame cube is shown in the Display window