OPERATION
Annex LSM 510
4-172 B 40-051 e 07/98
4.12.3 Detaching / Attaching the Scanning Module from / to Microscope Stands
Tool needed: 3 mm Allen key
☞
The user can remove the Scanning Module from one microscope and attach it to another
within a few minutes.
No adjustment
is required after the change-over. Described below is
the change-over from an Axioplan to an Axiovert 100 M in baseport configuration.
Before
the change-over,
shut down
the system as described in section 4.12 in order to avoid
damage to the system and loss of data.
Turn out both knurled-head screws (4-1/1) at the Scanning Module (4-1/2) fitted to the Axioplan.
Turn out M3 hexagon socket screw (4-1/3) with the Allen key.
Cautiously pull Scanning Module off the Axioplan stand.
Attach Scanning Module to the baseport of the Axiovert, minding the guide pins
(4-1/6), and secure it with the M3 hexagon socket screw (4-1/3).
Fasten Scanning Module to the baseport with two hexagon socket screws (4-1/5), using an offset
Allen key.
☞
As the Scanning Module is heavy, weighing about 14 kg, it is easier if the changeover is carried
out by two persons.
Pull off covering caps (4-1/4) from the CAN-BUS and RS232 interface ports at the rear of the Axiovert,
remove the two cables 457411-9011 (CAN-BUS) and 457411-9012 (RS 232) from the Axioplan, plug
them into the Axiovert and secure them there.
Switch the LSM 510 on with the REMOTE switch.
Click on the "Axiovert.ico" icon to update the system database with the new database of the
Axiovert 100 M microscope.
Restart the LSM program