
INTRODUCTION TO LASER SCANNING MICROSCOPY
LSM 510 Functions
B 40-051 e 07/98 5-25
1 - Wrap No normalization - the grey values are displayed modulo 256 (4096). If the result
is less than 0, the value 256 (4096) is added to it.
2 - Clip Negative values are set to 0.
3 - Normalize The resulting grey value range is scaled to the range 0...255 (0...4095).
4 - Shift/Clip 128 (2048) is added to the difference, then negative values are set to 0. Values
greater than 255 (4095) are set to 255 (4095).
Parameters:
Input 1 First input image sequence
Input 2 Second input image sequence
Output Output image sequence
Mode 1 - Wrap
2 - Clip
3 - Normalize
4 - Shift/Clip
Kommentare zu diesen Handbüchern