OPERATION
LSM 510 Image Acquisition (Acquire)
B 40-051 e 07/98 4-35
4.5.2.2 Epi-fluorescence observation (Axioplan 2
MOT
)
Turn on the HBO 100 W power supply with switch (4-37/2).
Push in the silver slider (4-37/8) on the microscope tube as far as it will go.
This opens the light path for specimen observation through the eyepieces.
Switch on the reflected light.
Pull out the occluding slider (4-37/1) to a light-passing position; actuate shutter switch (4-37/4) for
reflected light if it is in transmitted-light position.
To avoid excessive bleaching, expose the specimen to the minimum possible irradiation, i.e.
keep the irradiation time as short as possible.
In the Reflector Turret list box, select the reflector module (filter sets) to suit the type of fluorescence
excitation. Proceed as follows:
Open the Reflector Turret list box.
Click on the desired reflector module.
The reflector turret moves the selected
reflector module into the beam path.
The FITC filter set (FSET09) consists
of an excitation filter for the
450 - 490 nm spectral range, an
FT color splitter for 510 nm and an
LP long pass filter, which passes
emission light wavelengths greater
than 510 nm.
Fig. 4-38