1-22 06/99 B 40-015 e
1.8
Technical Data
Dimensions and weight
Dimensions (width × depth × height) ......................................................................180 × 400 × 430 mm
Footprint (recommended with mat)...................................................................................440 × 310 mm
Weight.............................................................................................................................. approx. 10 kg
Ambient conditions for
storage and transport (in packaging)
Permissible ambient temperature ....................................................................................... -40 to +50 °C
Permissible relative humidity ...................................................................................................max. 85 %
Atmospheric pressure ............................................................................................. 800 hPa to 1060 hPa
Operation
Permissible ambient temperature ........................................................................................ +5 to +40 °C
Permissible relative humidity ...................................................................................................max. 85 %
Altitude.............................................................................................................................. max. 2000 m
Atmospheric pressure ............................................................................................. 800 hPa to 1060 hPa
Pollution degree ................................................................................................................................... 2
Operating data
Category of use.................................................................................................................. closed rooms
Protection class.......................................................................................................................................I
Protection type................................................................................................................................IP 20
Electrical safety.................................................................in compliance with DIN EN 61010 (IEC 1010-1)
including CSA and UL directives
Excess voltage category .........................................................................................................................II
Radio interference suppression....................................................... in accordance with EN 55011, Class B
Insensitivity to noise....................................................................................in accordance with EN 50082
Line voltage....................................................................................................... 100 to 240 V AC ±10 %
....................................................................................................Change of line voltage is not required!
Line frequency...................................................................................................................... 50 to 60 Hz
Power consumption...............................................................................................................max. 80 VA
Fuses
for all line voltages above mentioned ............................................................. T 0.8 A; 250 V; 5 × 20 mm