3.2.2
Setting of Reflected-Light
Polarization
The Axiolab A microscope is ready for operation as
described in chapter 2, and switched on as
described in section 3.1.
• Pull out the plane glass pushrod (3-3/1).
• Pull out the dust protection sliders (3-3/2+6)
from the reflected-light illuminator H.
• Push polarization slider P (3-3/3) into the slider
mount (vertical slot) from the right.
• Push rotary analyzer slider A (3-3/4) in the upper
horizontal slider mount.
☞
For reasons of intensity, no filters
should be contained in the beam path.
• Set the analyzer slider to optimum contrast
(maximum dimming) in the field of view via the
adjustment lever.
☞
If possible, this analyzer position should not be changed.
• Reduce illumination aperture by turning the adjusting wheel (3-2/4) of the aperture diaphragm
downwards.
• To enhance the color contrast of anisotropic object details, the λ-sub compensator slider (3-3/5) can
be inserted in the lower horizontal slider mount.
The color intensity can then be varied via the lever on the λ-sub compensator slider.
3.2.3
Setting of Transmitted-Light Polarization with extended Polarization Equipment
The optional, extended polarization equipment includes the following items:
-
rotary Pol stage (3-4/3)
-
Pol object guide (3-4/4)
-
swing-in polarizer D (3-4/5)