Carl Zeiss Axiolab A
4-6 B 40-015 e 06/99
(3) Changing the 6 V, 30 W Transmitted-
Light Halogen Lamp
The following procedure is required to change the
lamp of the integrated 6 V, 30 W illuminator:
Switch off the extern lamp supply; if required,
allow the illuminator to cool down for approx.
15 minutes.
Disconnect the line cable from the line.
Hold the ventilation grille (4-3/1) on its grip and
pull it off in upward direction.
Remove the lamp insert by pressing its handle
(4-3/2) together.
Replace the halogen lamp 6 V, 30 W (4-3/3).
Do not touch the lamp bulb with your bare hands; if required, clean the bulb using pure
alcohol before switching it on for the first time to prevent dirt from burning in.
Make sure that the lamp carrier plate exactly fits the positioning pins.
Attach lamp insert again by slightly pressing the handle together
Attach ventilation grille again until stop.
4.3
Requesting Service
All repairs of mechanical, optical or electronic components inside the instrument and of the electrical
components of the Axiolab A may only be performed by Carl Zeiss service staff or specially authorized
personnel.
To ensure the optimum setting and trouble-free function of your microscope even for a longer period of
time, we would recommend you to conclude a service/maintenance contract with Carl Zeiss.
In the case of subsequent orders or when service is required, please get in touch with your local Carl
Zeiss agency.
Fig. 4-3 Changing the 6 V, 30 W
transmitted-light halogen lamp