Axiolab A Carl Zeiss
B 40-015 e 06/99 4-5
(2) Changing the 6 V, 30 W Reflected-Light
Halogen Lamp
The following procedure is required to change the
lamp of the 6 V, 30 W reflected-light halogen
illuminator:
Switch off the lamp supply; if required, allow
the illuminator to cool down for approx. 15
minutes.
Disconnect the line cable from the line.
Remove the connection cable of the reflected-
light illuminator from the socket at the rear of
the instrument.
Turn the cover (4-2/3) to the left and remove it
in backward direction.
To facilitate the lamp change, loosen the
hexagonal screw (4-2/1) by a few turns and
remove the mount (4-2/2) from the reflected-
light illuminator (the tools are listed together
with the accessories)
Loosen both fixation screws (4-2/5) by turning them to the left and remove the carrier plate (4-2/4)
containing the lamp.
Remove new 6 V, 30 W halogen lamp from the packaging and insert it into the mount (4-2/2); one
tip of the ceramic base must engage in the centering notch of the carrier plate, and the carrier plate
must fit well to the second tip.
Do not touch the lamp bulb with your bare hands; if required, clean the bulb using pure
alcohol before switching it on for the first time to prevent dirt from burning in.
Attach reflected-light illuminator again in such a way that the halogen lamp is positioned vertically.
Tighten hexagonal screw (4-2/1).
Attach cover (4-2/3) and lock it by turning it to the right.
Connect the cable of the reflected-light illuminator to the rear of the microscope stand.
Fig. 4-2 Changing the 6 V, 30 W reflected-
light halogen illuminator