Axiolab A Carl Zeiss
B 40-015 e 06/99 4-5
(2) Changing the 6 V, 30 W Reflected-Light
Halogen Lamp
The following procedure is required to change the
lamp of the 6 V, 30 W reflected-light halogen
illuminator:
• Switch off the lamp supply; if required, allow
the illuminator to cool down for approx. 15
minutes.
• Disconnect the line cable from the line.
• Remove the connection cable of the reflected-
light illuminator from the socket at the rear of
the instrument.
• Turn the cover (4-2/3) to the left and remove it
in backward direction.
• To facilitate the lamp change, loosen the
hexagonal screw (4-2/1) by a few turns and
remove the mount (4-2/2) from the reflected-
light illuminator (the tools are listed together
with the accessories)
• Loosen both fixation screws (4-2/5) by turning them to the left and remove the carrier plate (4-2/4)
containing the lamp.
• Remove new 6 V, 30 W halogen lamp from the packaging and insert it into the mount (4-2/2); one
tip of the ceramic base must engage in the centering notch of the carrier plate, and the carrier plate
must fit well to the second tip.
☞
Do not touch the lamp bulb with your bare hands; if required, clean the bulb using pure
alcohol before switching it on for the first time to prevent dirt from burning in.
• Attach reflected-light illuminator again in such a way that the halogen lamp is positioned vertically.
Tighten hexagonal screw (4-2/1).
• Attach cover (4-2/3) and lock it by turning it to the right.
• Connect the cable of the reflected-light illuminator to the rear of the microscope stand.
Fig. 4-2 Changing the 6 V, 30 W reflected-
light halogen illuminator