Setting of Transmitted-Light Brightfield (KÖHLER Illumination)
Make the microscope ready for operation as
described in chapter 2 and section 3.1.
• Switch on the illumination, place a high-
contrast specimen on the stage, and focus
using the 10× objective.
• Close luminous-field diaphragm (3-10/3) via the
knurled wheel.
• Use drive (3-10/5) to move ABBE condenser
close to the upper stop position until the image
of the luminous-field diaphragm appears in
focus.
• Use the centering screws (3-10/2) to move the
luminous-field diaphragm to the center of the
field of view.
• Open the luminous-field diaphragm until its
edge just disappears from the field of view.
• Control the contrast via the aperture diaphragm
(3-10/4), which should be open by approx. 2/3
of its size (marking on condenser positioned to
0.5).
For further information on the use of the
Axiolab A for transmitted-light examinations and
the relevant accessories please see the manual
Axiolab - Transmitted Light and Epi-Fluorescence
(G 42-130/II e).