Zeiss Axiolab A Bedienerhandbuch Seite 56

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 76
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 55
Carl Zeiss Axiolab A
3-14 B 40-015 e 06/99
3.2.5
Setting of Transmitted-Light Brightfield (KÖHLER Illumination)
Make the microscope ready for operation as
described in chapter 2 and section 3.1.
Switch on the illumination, place a high-
contrast specimen on the stage, and focus
using the 10× objective.
Close luminous-field diaphragm (3-10/3) via the
knurled wheel.
Use drive (3-10/5) to move ABBE condenser
close to the upper stop position until the image
of the luminous-field diaphragm appears in
focus.
Use the centering screws (3-10/2) to move the
luminous-field diaphragm to the center of the
field of view.
Open the luminous-field diaphragm until its
edge just disappears from the field of view.
Control the contrast via the aperture diaphragm
(3-10/4), which should be open by approx. 2/3
of its size (marking on condenser positioned to
0.5).
For further information on the use of the
Axiolab A for transmitted-light examinations and
the relevant accessories please see the manual
Axiolab - Transmitted Light and Epi-Fluorescence
(G 42-130/II e).
Fig. 3-10 Setting of transmitted-light
brightfield
Seitenansicht 55
1 2 ... 51 52 53 54 55 56 57 58 59 60 61 ... 75 76

Kommentare zu diesen Handbüchern

Keine Kommentare