Carl Zeiss Axiolab A
3-6 B 40-015 e 06/99
(1) Centering of Objectives (only for Axiolab Pol catalogue No. 450910-0000-000)
Fig. 3-4 Centering of objectives
When the Pol stage (3-4/3) and uncentered objectives are turned, object details above the center of the
eyepiece cross migrate to circular paths (3-4/1 - dashed line). The following procedure is required to
center the objectives:
Unlock the stage stop by turning screw (3-4/6) by approx. ¼ rotation in counterclockwise direction.
Swing in transmitted-light objective of medium magnification (20× - 50×) via the nosepiece.
Insert SW 1.5 key (3-4/2) in the drilled centering holes on the nosepiece.
Turn the specimen stage and search for the apparent center of rotation (cross (3-4/1)). The center of
rotation is also the center of the stage rotation.
Use the keys to move this center of rotation to the intersecting point of the eyepiece crosslines and
therefore into the optical axis of the microscope.
Repeat the procedure with a distinct, small object detail in the direct proximity of the center of the
rotary stage.
Carefully remove the key.
Swing in the neighboring objective; search for the center of rotation and the distinct object details
and perform the adjustment procedure as described above.
To retain the centering status, it is absolutely necessary to change the objectives only via the
knurled ring of the objective nosepiece.