Axiolab A Carl Zeiss
B 40-015 e 06/99 1-23
Mechanical data of specimen stages
Stage models............................................................................. fixed stage/glide stage/mechanical stage
Maximum specimen height:
stage at bottom stop (not possible for mechanical stage)................................................... 45 mm
stage at top stop .............................................................................................................. 20 mm
Light source
6 V, 30 W halogen illuminator:
halogen lamp with square flat-core filament .............................................. HLW S 5-A, 6 V, 30 W
lamp voltage (variable).................................................................................... >1.5 V to max. 6 V
power .................................................................................................................................30 W
color temperature at end stop .......................................................................................... 3200 K
light flux at 6 V................................................................................................................. 510 lm
average life at 6 V/4.8 V ......................................................................................... 300 h/2000 h
illuminated area...................................................................................................... 1.7 × 1.7 mm
Opto-mechanical data
Stage focusing.................................................................................................... coaxial coarse/fine drive
lifting range: 20 mm
coarse drive: 1 rotation = 4 mm
fine drive: 1:10 step-down ratio
Objective change ................................................................................. manually via 5-position nosepiece
Maximum field of view ................................................................................................................ 20 mm
Maximum object field .................................................................................................................... 8 mm
Binocular 30°/20 H reflected-light tube with camera/TV port:
maximum field number............................................................................................................20
interpupillary distance.................................................................variable between 55 and 75 mm
viewing angle.........................................................................................................................30°
viewing height................................................................................................................ 440 mm
observation and camera/video port ........................................................................ tube factor 1×
observation and camera/video port ........................................................ 100%/0% or 0%/100%
camera/video port .............................................................................................. 60 mm interface
Eyepieces................................................................................... E-PL 10×/20 Br. and E-PL 10×/20 Br. foc.
plug-in diameter............................................................................................................... 30 mm
Kommentare zu diesen Handbüchern