Troubleshooting
Troubleshooting on the Axiolab A microscope is limited to only a few actions:
Checking the line cable
Checking the set instrument voltage and changing the fuses
Checking the illumination and changing the lamp
Defective fuse(s) change the fuses according to paragraph (1)
defective halogen lamp change the lamp according to paragraph (2) or (3)
(1) Check the Line Cable and Fuses and
Change Fuses
Make sure to disconnect the instru-
ment from the line first!
Check line cable (4-1/1) and replace it, if
required.
Use 2.0 mm watchmaker’s screwdriver to press
the two springs at the side of the fuse holder
(4-1/2) to the inside via the groove and remove
fuse holder.
Check fuse inserts (4-1/3) and replace defective
fuses with new ones. Make sure to use only
T 0.8 A; 250 V; 5 × 20 mm fuses!
Insert fuse holder again until the springs at the
side can be heard to engage.
Connect the instrument to the line.
Fig. 4-1 Changing the fuses